CLINICAL CHEMISTRY | SCIENTIFIC ARTICLES
Quality Control and Calibration in HDL Cholesterol Testing: Ensuring Analytical Accuracy and Consistency
Explore HDL cholesterol quality control, calibration, two-level controls, reagent stability and practical approaches to consistent clinical chemistry results.
1. Why Is Quality Assurance Important in HDL-C Testing?
HDL cholesterol (HDL-C) is a routine component of the lipid profile.
Reliable clinical interpretation depends on analytical results that
remain consistent over time and are suitable for their intended use.
Even a well-designed enzymatic method requires appropriate calibration,
internal quality control (IQC) and documented verification on the
analyzer used by the laboratory.
Quality assurance is not limited to a single test performed during
instrument installation. It is a continuous process that includes
preventive checks, routine monitoring, investigation of nonconformities
and corrective actions.
This article discusses automated direct HDL-C methods and explains
the materials specified in the instructions for use (IFU) of
Archem Diagnostics Direct HDL Cholesterol.
2. Calibration: Establishing the Measurement Relationship
Calibration establishes the relationship between instrument response
and an assigned analyte concentration. For HDL-C, calibration should
follow reagent-specific application parameters, calibrator instructions
and analyzer software requirements.
A stated calibration stability interval is conditional and does not
mean that quality-control findings can be disregarded during that period.
Reagent lot changes, major maintenance procedures, shifts in control
results or unacceptable analytical performance may require
recalibration or further investigation.
Calibrators and control materials serve different purposes:
a calibrator establishes the measurement relationship,
whereas a control material monitors whether the analytical
process remains under control.
The assigned values and metrological traceability of the calibrator
should be evaluated using the relevant product documentation.
3. Internal Quality Control: Two Levels and Routine Monitoring
Internal quality control monitors routine method performance using
materials with expected analyte concentrations.
Using two different control levels allows laboratories to assess
performance at more than one concentration and may help identify
problems that could be missed with a single control level.
The Archem Direct HDL-C IFU specifies that at least two control
levels should be tested once every 24 hours.
Each laboratory must also establish its own quality-control plan
and acceptance criteria. Additional controls may be appropriate
after maintenance or reagent changes.
Relevant control results should be evaluated before patient
results are released.
Time-based monitoring tools, such as Levey–Jennings charts,
can help laboratories identify trends, shifts and outlying
control values.
4. Interpreting Precision, Bias and Total Analytical Error
Precision describes the closeness of repeated measurements
under specified conditions. Bias represents systematic
deviation from a reference or comparison target.
Both should be considered when establishing analytical
performance requirements.
A low coefficient of variation (CV) alone does not demonstrate
sufficient accuracy. Similarly, a high correlation coefficient
does not establish complete agreement between methods.
Performance acceptance criteria should be defined before
the analytical data are interpreted.
Under the evaluation conditions described in the Archem HDL-C IFU,
within-run CV values of 0.60–1.01% and
day-to-day CV values of 2.05–2.18%
were reported.
These figures represent manufacturer study results and
do not replace the laboratory's own performance verification.
5. Factors That May Affect HDL-C Results
Sample-handling errors, reagent temperature, inadequate mixing,
pipetting problems, optical-system performance, carry-over,
reagent deterioration and inappropriate calibration settings
may contribute to analytical variation.
Analyzer cooling capacity is particularly important when
assessing on-board reagent stability.
Potential interferences and specimen characteristics
should be evaluated in conjunction with the current IFU.
When unexpected results occur, laboratories should investigate
sample integrity and instrument alerts rather than relying
solely on repeat testing.
Analyzer applications may differ in sample and reagent volumes,
measurement wavelengths, reaction timing and wash parameters.
The inclusion of an instrument in a corporate validation list
does not necessarily mean that every assay has been validated
on that particular model.
6. Archem HDL-C: Calibrator, Controls and Stability
Archem Direct HDL Cholesterol is a ready-to-use liquid,
two-reagent homogeneous enzymatic colorimetric assay
for quantitative measurement in human serum and plasma.
For calibration, the IFU specifies the lyophilized
HDL-LDL (Arcal Lipid) Calibrator.
For internal quality control, the IFU identifies the
lyophilized Arcon N (Level 1) and
Arcon P (Level 2) control materials.
Commercial control materials with values established
for this method may also be used when consistent
with the IFU requirements.
The IFU reports 30 days of calibration stability,
depending on the analyzer application and cooling capacity.
Opened reagent bottles remain stable for
30 days at 2–8°C under optimum conditions.
On-board stability depends on cooling performance
and carry-over characteristics.
Reagents must be stored according to the instructions
and must not be frozen.
Archem Direct HDL-C: Key Quality Parameters
|
Calibrator
|
HDL-LDL (Arcal Lipid), lyophilized
|
|
Level 1 Control
|
Arcon N, lyophilized
|
|
Level 2 Control
|
Arcon P, lyophilized
|
|
Control Frequency
|
At least two levels once every 24 hours
|
|
Calibration Stability
|
30 days; dependent on analyzer application and cooling
|
|
Open-Vial Stability
|
30 days at 2–8°C under optimum conditions
|
|
Within-Run CV
|
0.60–1.01% (manufacturer study)
|
|
Day-to-Day CV
|
2.05–2.18% (manufacturer study)
|
7. When Control Results Exceed Acceptance Limits
When a control result exceeds its acceptance limits,
the relevant analytical process should be evaluated
before patient results are released.
Repeatedly measuring the control without investigating
the underlying cause until an acceptable value appears
is not an appropriate approach.
Laboratories may review control preparation and storage,
reagent expiration dates and lot numbers, maintenance history,
pipetting performance, temperature conditions and
calibration status.
The Archem IFU instructs laboratories to perform calibration
when quality-control results fall outside the acceptance limits.
Following corrective action, appropriate controls should
be tested again. The identified cause, actions taken and
potential impact on previously reported results should
be documented.
8. Establishing a Sustainable Quality Management Workflow
An effective quality-control plan defines control frequency,
concentration levels, target values, acceptance limits,
review responsibilities and procedures for managing
nonconforming results.
Procedures following reagent lot changes, calibration
and instrument maintenance should also be established.
Regular review of control trends and participation
in suitable external quality assessment programs
can complement daily internal quality control.
These activities do not replace verification of the
reagent–analyzer application before clinical use.
Consistent results in direct HDL-C measurement depend
on the combined implementation of appropriate procedures,
suitable materials, instrument maintenance, trained
personnel and documented review.
Conclusion
Calibration establishes the analytical response,
while internal quality control monitors whether
that response remains within acceptable limits.
When combined with application verification
and documented corrective actions, these processes
support the consistency of HDL-C results.
For its Direct HDL-C method, Archem Diagnostics specifies
the HDL-LDL (Arcal Lipid) Calibrator
and Arcon N/P controls.
Laboratories should follow current product instructions
and applicable quality requirements.
Selected References
-
Archem Diagnostics.
Direct HDL Cholesterol (D-HDL-C), Instructions for Use.
Rev. V3.5, August 2023.
-
Clinical and Laboratory Standards Institute (CLSI).
Evaluation of Precision of Quantitative Measurement Procedures.
EP05-A3, 2014.
-
Clinical and Laboratory Standards Institute (CLSI).
Statistical Quality Control for Quantitative Measurement Procedures:
Principles and Definitions.
C24, 4th ed., 2016.
-
Clinical and Laboratory Standards Institute (CLSI).
Evaluation of Stability of In Vitro Diagnostic Reagents.
EP25-A, 2009.
-
Clinical and Laboratory Standards Institute (CLSI).
Interference Testing in Clinical Chemistry.
EP07, 3rd ed., 2018.