Article

HbA1c Testing on Automated Clinical Chemistry Analyzers: Workflow, Performance and Key Considerations

HbA1c Testing on Automated Clinical Chemistry Analyzers: Workflow, Performance and Key Considerations

HbA1c Testing on Automated Clinical Chemistry Analyzers: Workflow, Performance and Key Considerations

HbA1c testing is traditionally associated with dedicated analytical systems, but immunoturbidimetric assays also make it possible to perform HbA1c measurement on compatible automated clinical chemistry analyzers. This approach allows laboratories to integrate HbA1c into established chemistry workflows while maintaining defined calibration, quality-control and analytical performance requirements.

For laboratories evaluating an HbA1c reagent for an automated chemistry analyzer, the analytical principle is only one part of the decision. Analyzer-specific application performance, sample preparation, reagent handling, calibration stability, quality control, measuring interval, precision, interference and standardization should all be considered as part of the complete analytical system.

Can HbA1c Be Measured on a Clinical Chemistry Analyzer?

Yes. HbA1c can be measured on compatible automated clinical chemistry analyzers using appropriately developed and validated assay applications.

Immunoturbidimetric HbA1c assays combine an antibody-based reaction with photometric detection. This analytical format allows the assay to be adapted to the photometric systems already available on many automated chemistry platforms.

Key concept: an immunoturbidimetric HbA1c assay can allow laboratories to integrate HbA1c testing into an existing automated clinical chemistry workflow without necessarily requiring a separate dedicated HbA1c analyzer.

The exact application parameters, calibration procedures and performance characteristics remain analyzer-specific and should therefore be validated for the intended analytical platform.

From Whole Blood Sample to HbA1c Result: The Laboratory Workflow

HbA1c measurement on a clinical chemistry analyzer begins with appropriate handling of the whole blood specimen and continues through sample preparation, automated reaction, photometric measurement and result calculation.

1 Whole Blood Sample
2 Sample Mixing
3 Hemolysis
4 Analyzer Loading
5 Automated Reaction
6 Photometric Measurement
7 HbA1c Result

1. Whole Blood Sample Preparation

HbA1c is measured from whole blood because glycated hemoglobin is located within erythrocytes. For Archem HbA1c II DIRECT, anticoagulated human venous whole blood is the recommended specimen type, with K2-EDTA or K3-EDTA specified as recommended anticoagulants.

Samples should be mixed carefully before testing to ensure homogeneous distribution of erythrocytes.

2. Hemolysate Preparation

Erythrocytes are lysed before measurement so that hemoglobin becomes available for the analytical reaction.

In the Archem HbA1c II DIRECT procedure, whole blood is mixed with the specified lysing solution and incubated to prepare the hemolysate before analysis.

3. Automated Immunoturbidimetric Reaction

Once prepared, the sample is processed on the chemistry analyzer using the specified application. The assay uses an HbA1c-specific immunoturbidimetric reaction, with the resulting change in turbidity measured photometrically.

For a detailed explanation of latex interaction, antibody binding, agglutination and photometric detection, see our article “How Does Immunoturbidimetric HbA1c Testing Work?”

4. Result Calculation and Reporting

HbA1c results may be reported according to the NGSP system as a percentage or according to the IFCC system in mmol/mol. Appropriate calibration and standardization support comparability of results within the defined analytical system.

Why Integrate HbA1c into a Routine Chemistry Workflow?

The possibility of performing HbA1c testing on an established chemistry platform may offer operational advantages for laboratories that already use automated clinical chemistry systems.

Existing Analyzer Infrastructure HbA1c testing can be integrated into compatible chemistry platforms already used for routine laboratory testing.
Automated Photometric Measurement The immunoturbidimetric reaction can be processed using the analyzer's automated photometric workflow.
Defined Calibration Procedures Analyzer-specific applications can incorporate dedicated HbA1c calibration procedures.
Routine Quality Control HbA1c controls can be incorporated into established laboratory quality-control procedures.
Liquid Reagent Format Ready-to-use liquid reagents can support routine laboratory reagent handling.
Multi-Platform Applications A validated reagent system can be applied across multiple compatible chemistry analyzer platforms.

Reagent Format, Storage and Handling

Reagent format and stability influence both analytical performance and routine laboratory workflow.

Archem HbA1c II DIRECT is supplied as a liquid, dual-reagent system. The reagents are ready to use and should be stored at 2–8°C.

Opened reagent vials are stable for 30 days at 2–8°C under optimum conditions. However, on-board stability is strongly related to the cooling specifications and carry-over characteristics of the autoanalyzer.

On-board reagent stability should not be considered independently from the analyzer. Cooling performance, reagent compartment conditions and carry-over characteristics can influence practical on-board stability.

Calibration and Quality Control

Calibration and quality control are essential components of an HbA1c analytical system.

HbA1c Calibration

Archem HbA1c II DIRECT requires a 4-level HbA1c calibrator. Calibration stability is stated as 30 days, depending on the characteristics of the analyzer application and the cooling capacity of the instrument.

Analyzer-specific calibrator references are available for selected platforms, supporting application-specific implementation.

HbA1c Quality Control

Low and high HbA1c control materials are used to monitor analytical performance. According to the current Archem Instructions for Use, two control levels should be run once every 24 hours.

Each laboratory should establish its own quality-control procedures according to its regulatory requirements, laboratory policy and analytical performance goals.

Analytical Performance: What Should Laboratories Evaluate?

When selecting an HbA1c reagent for an automated chemistry analyzer, analytical performance should be considered together with analyzer compatibility and workflow requirements.

Performance Characteristic Archem HbA1c II DIRECT
Measuring Interval 4.0–15.5% HbA1c
IFCC Measuring Interval 20–140 mmol/mol
Limit of Detection 2.7%
Limit of Quantitation 4.0%
Within-Run CV at 5.46% HbA1c 1.45%
Within-Run CV at 10.10% HbA1c 1.73%
Day-to-Day CV at 5.46% HbA1c 2.81%
Day-to-Day CV at 10.10% HbA1c 2.72%
Total Bilirubin Tested Up to 48 mg/dL
Triglycerides Tested Up to 2000 mg/dL

Performance characteristics are established under defined analytical conditions. Results may vary slightly according to analyzer, application configuration and laboratory conditions.

Validated HbA1c Applications Across Clinical Chemistry Platforms

Archem HbA1c II DIRECT has been validated across a broad portfolio of automated clinical chemistry analyzers, supporting laboratories and distributors working with different analyzer platforms and workflow requirements.

The validated application portfolio covers systems from leading global and regional clinical chemistry analyzer manufacturers.

Manufacturer / Platform Validated Analyzer Models Status
Beckman / Olympus AU400, AU600, AU480, AU680, AU700, AU5400 Validated
Beckman / Olympus AU5800, DxC 700 AU Validated NGSP Certified*
Mindray BS300, BS360, BS380, BS400, BS430 Validated
Mindray BS120, BS200, BS220, BS240, BS260 Validated
Mindray BS800, BS2000 Validated NGSP Certified*
Roche cobas INTEGRA 400 plus, c 111, c 303, c 311, c 501, c 502, c 503, c 701, c 702, c 703 Validated
Dirui CS-6400, CS-4000, CS-1600, CS-1300B, CS-1200, CS-600B, CS-T240 Validated
Erba Mannheim XL200, XL600, XL640, XL1000 Validated
Snibe Biossays 240 Plus, C8, C10 Validated
YHLO iBC 900, iBC 2000 Validated
Thermo Fisher Scientific Konelab 20, Konelab 30, Konelab 60, Indiko, Indiko Plus Validated
Medical System MS-380, MS-680 Validated
Rayto Chemray 120, 240, 330, 380, 420, 2000 Validated
Prestige / Tokyo Boeki Biolis 24i, 30i, 50i, Labmax 240 Validated
Edif / Biolabo CORE, ALCOR, SPHERA, DUO, ORION 240, PULSAR 250, ORION 320, Kenza 240 TX, Kenza 450 TX, Kenza One Validated
URIT CA-200, CA-331A, CA-430A, CA-640A, CA-810B Validated
Biotecnica Instruments BT1500, BT3000, BT3000 Plus, BT3500, BT3600 Validated
Bioelab AS-160, AS-280, AS-480, AS-600, AS-800 Validated
Biolyzer / Randox / Furuno / Sysmex Biolyzer 300, Randox Daytona, Furuno CA-180, CA-270, CA-310, Sysmex BX-3010, BX-4000 Validated
Biolyzer / Furuno / Randox Biolyzer 600, Furuno CA-400, Randox Imola Validated
Biobase BK-200, BK-280, BK-400, BK-600, BK-1200, BK-2000 Validated
Biosystems A15, A25, BA200, BA400 Validated
TECO Coralyzer Junior, Coralyzer Smart, TC-160, TC-240, TC-480, TC-600, TC-800 Validated
HumaStar 300, 600 Validated
Hitachi 704, 717, 747, 902, 911, 912, 917, 922 Validated
Instrumentation Laboratory ILab 300, ILab 300 Plus, ILab 500, ILab 600, ILab 900 Validated
Selectra 2E, Flexor, Pro S, Pro M Validated
Agappe Mispa CX4, Mispa CXL Pro, Mispa FAB120 Validated
HORIBA Pentra C200, Pentra C400, Yumizen C230, C240, C560 Validated
Gesan Chem 200 Pro, Chem 300 Plus, Chem 400 Pro Validated
Genrui GS100, GS480A Validated
Landwind LW C100, LW C100 Plus, LW C200E, LW C400, LW C480 Validated
Autobio AutoChem B861, AutoChem B2000 Validated
Abbott ARCHITECT c4000, c8000, c16000 Validated NGSP Certified on selected systems*
Abbott Alinity c Series Validated NGSP Certified*
Siemens Healthineers Dimension RxL Max, Dimension XPAND, Atellica, EXL 200 Validated
Siemens Healthineers ADVIA 1200, ADVIA 1650, ADVIA 1800, ADVIA 2400, ADVIA XPT Series Validated
Siemens Healthineers Atellica CH Series Validated
Validity aV-280, aV-500 Validated

Analyzer-specific applications are validated under defined Archem validation conditions. Instrument and manufacturer names are used for identification purposes only and remain trademarks of their respective owners.

NGSP-Certified HbA1c Systems

In addition to analyzer application validation, Archem has completed NGSP Manufacturer Certification for specified HbA1c analytical systems.

For the certification period May 1, 2026 – May 1, 2027, certified configurations include the following platforms:

Manufacturer NGSP-Certified Platform
Abbott Alinity C
Abbott ARCHITECT c4000
Abbott ARCHITECT c8000
Beckman Coulter DxC 700 AU
Olympus AU480
Olympus AU5800
Mindray BS2000M
Mindray BS800 / BS800M

*NGSP manufacturer certification applies to the specific analytical system, reagent/calibrator configuration and instrumentation evaluated during certification. NGSP certification should therefore not be interpreted as a blanket certification for all analyzer applications.

What Should Laboratories Consider When Choosing an HbA1c Reagent?

An HbA1c reagent should be evaluated as part of a complete analytical system rather than as an isolated reagent.

Analyzer Compatibility Confirm that an appropriate analyzer-specific application is available.
Analytical Principle Understand how HbA1c is measured and how the assay fits the laboratory workflow.
Standardization and Traceability Review NGSP and IFCC-related standardization information where applicable.
Precision Assess repeatability and within-laboratory performance at clinically relevant concentrations.
Measuring Interval Confirm that the assay covers the clinically relevant HbA1c range.
Interference Data Review available information regarding potentially interfering substances.
Calibration and QC Consider calibration frequency, control requirements and material availability.
Reagent Stability Evaluate storage, open-vial stability and analyzer-dependent on-board conditions.

Archem HbA1c II DIRECT immunoturbidimetric reagent for automated clinical chemistry analyzers

Archem HbA1c II DIRECT

HbA1c II DIRECT is a liquid, dual-reagent immunoturbidimetric HbA1c assay developed for quantitative HbA1c determination in human whole blood on compatible automated clinical chemistry analyzers.

Method Immunoturbidimetric
Sample Anticoagulated human venous whole blood
Reagent Format Liquid, dual reagent
Preparation Ready to use
Measuring Interval 4.0–15.5% / 20–140 mmol/mol
Calibration Stability 30 days*
Opened-Vial Stability 30 days at 2–8°C under optimum conditions
Reporting NGSP (%) / IFCC (mmol/mol)

*Calibration stability depends on analyzer application characteristics and cooling capacity. On-board reagent stability is analyzer-dependent.

Frequently Asked Questions

Can HbA1c be tested on an automated clinical chemistry analyzer?

Yes. Immunoturbidimetric HbA1c assays can be implemented on compatible automated clinical chemistry analyzers using validated analyzer-specific applications.

Does immunoturbidimetric HbA1c testing require a dedicated HbA1c analyzer?

Not necessarily. A validated immunoturbidimetric HbA1c application can allow HbA1c testing to be integrated into a compatible routine clinical chemistry analyzer.

What sample is used for HbA1c testing on a chemistry analyzer?

HbA1c testing is performed using whole blood. For Archem HbA1c II DIRECT, anticoagulated human venous whole blood collected with K2-EDTA or K3-EDTA is recommended.

Which analyzers are validated for Archem HbA1c II DIRECT?

Archem has validated HbA1c applications across a broad range of clinical chemistry analyzer platforms including systems from Beckman Coulter/Olympus, Abbott, Mindray, Roche, Siemens Healthineers, Erba Mannheim, Dirui, Snibe, YHLO, Thermo Fisher Scientific, Randox, HORIBA, Hitachi and many other manufacturers.

What should a laboratory consider when selecting an HbA1c reagent?

Laboratories should consider analyzer compatibility, analytical performance, measuring interval, precision, interference data, calibration stability, quality-control requirements, reagent stability and standardization or traceability information.

Conclusion

Immunoturbidimetric HbA1c testing makes it possible to integrate HbA1c measurement into compatible automated clinical chemistry workflows.

Successful implementation depends not only on the reagent itself, but on the complete analytical system: sample preparation, analyzer application, calibration, quality control, analytical performance, reagent stability and standardization.

Archem HbA1c II DIRECT combines a ready-to-use liquid reagent system with a broad portfolio of validated analyzer applications, defined analytical performance characteristics and NGSP-certified configurations for selected automated clinical chemistry platforms.

References

  1. Archem Diagnostics. HbA1c II DIRECT Instructions for Use. Rev. V3.6, July 2026.
  2. Clinical and Laboratory Standards Institute (CLSI). Evaluation of Stability of In Vitro Diagnostic Reagents; EP25.
  3. Clinical and Laboratory Standards Institute (CLSI). Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; EP17.
  4. Clinical and Laboratory Standards Institute (CLSI). Evaluation of Linearity of Quantitative Measurement Procedures; EP06.
  5. Clinical and Laboratory Standards Institute (CLSI). Evaluation of Precision of Quantitative Measurement Procedures; EP05.
  6. Clinical and Laboratory Standards Institute (CLSI). Interference Testing in Clinical Chemistry; EP07.
  7. NGSP. National Glycohemoglobin Standardization Program. Manufacturer certification and HbA1c standardization resources.